The DP120 differential voltage probe is designed to make measurements on signals with floating reference. The DP120 has two probes with shrouded banana probe tips (one red, one black) and shielded cable for improved noise immunity. Input impedance between each input to shielding BNC is 5 MΩ in parallel with 6 pF. Input impedance between the inputs is 10 MΩ with a capacitance of 5 pF.
Input: Probes with shrouded banana probe tip
Probe length: 1.5 meter (60 inch)
Maximum Input Voltage to Ground: 600V rms CAT III per IEC 1010-2-031 / 1000Vrms CAT II, per IEC 1010-2-031
Maximum Differential Voltage: 1000V DC or1000Vrms or 1200V (DC + ACpeak)
Output Cable: Safety designed 0.5m (5ft) BNC Cable